Sequential Injection Titration with Spectrophotometric Detection for the Assay of Acidity in Fruit Juices
Jaroon JAKMUNEEThitima RUJIRALAIKate GRUDPAN
著者情報
ジャーナル フリー

2006 年 22 巻 1 号 p. 157-160

詳細
抄録
A simple sequential injection analysis (SIA) with spectrophotometric detection for an assay of acidity in fruit juice was investigated. An alkaline reagent (sodium hydroxide), a sample and an indicator (phenolphthalein) were first aspirated and stacked as adjacent zones in a holding coil. With flow reversal through a reaction coil to the detector, zone penetration occurred, leading to a neutralization reaction that caused a decrease in the color intensity of the indicator being monitored for absorbance at 552 nm. The effects of various parameters were studied. Linear calibration graphs for acidities of 0.2 - 1.0 and 0.5 - 2.5% w/v citric acid as a standard, with a relative standard deviation of 1% (acidity of 0.3 - 0.6% w/v as citric acid, n = 11) and a sample throughput of 30 samples h-1, were achieved. The developed method was validated by a standard titrimetric method for assaying the acidity of fruit juice samples.
著者関連情報
© 2006 by The Japan Society for Analytical Chemistry
前の記事 次の記事