Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Fully Validated Ultra-performance Liquid Chromatography–Tandem Mass Spectrometry Method for the Determination of Fimasartan in Human Plasma
Ji Yeon HYUNHyun-Ju KIMMi-Ri GWONHae Won LEESook Jin SEONGYoung-Ran YOON
著者情報
ジャーナル フリー

2015 年 31 巻 12 号 p. 1335-1339

詳細
抄録
Fimasartan is a novel angiotensin II receptor blocker with strong anti-hypertensive activity. In this study, a more rapid and sensitive ultra-performance liquid chromatography–tandem mass spectrometry (UPLC-MS/MS) method to determine fimasartan in human plasma was developed and fully validated. The quantification of analytes was conducted by MS/MS in a multiple reaction monitoring mode at m/z 502.2 → 207.1 for fimasartan and m/z 526.3 → 207.1 for the internal standard (IS, BR-A-563). The method showed a linear response from 3 to 1000 ng/mL (r > 0.9950). The intra- and inter-day accuracy values were 86.9 – 98.2 and 93.3 – 100.1%, respectively. The intra- and inter-day precision values were 2.0 – 3.1 and 0.8 – 8.0%, respectively. This UPLC-MS/MS method was validated for specificity, linearity, precision, accuracy, recovery, system suitability, and stability, and was found to be acceptable for bioanalytical applications. Finally, this fully validated method was successfully applied to a pharmacokinetic study of fimasartan in healthy volunteers following oral administration.
Fullsize Image
著者関連情報
© 2015 by The Japan Society for Analytical Chemistry
前の記事 次の記事
feedback
Top