Subacute Presentation of Right Ventricular Perforation after Pacemaker Implantation
Sachi KoyamaKeiichi ItataniShunei KyoRie AoyamaTaizo IshiyamaKazumasa HaradaMinoru Ono
著者情報
ジャーナル フリー

2013 年 19 巻 1 号 p. 73-75

詳細
抄録
We report an 87-year-old woman with right ventricular perforation due to a permanent pacemaker lead detected 4-days after implantation. The pacemaker lead was seen to perforate through the myocardium and pericardium and to reach the left pleural cavity. We removed the wire surgically by median sternotomy. The pericardial effusion was cloudy and yellowish, suggesting infection. However, no bacteria were detected by bacterial cultures of the pericardial effusion and pacing wire. The patient developed neither mediastinitis nor sepsis after the operation, and a new pacemaker was implanted safely one month later.