Morphological Evaluation of Proximal Anastomosis by PAS-Port® System in Patients with Long-Term Patent Grafts
Muneyasu KawasakiTakeshiro FujiiMasanori HaraYuki SasakiTomoyuki KatayanagiShinnosuke OkumaYoshinori Watanabe
Author information
JOURNALS FREE ACCESS

Volume 21 (2015) Issue 2 Pages 172-177

Details
Download PDF (1324K) Contact us
Abstract

Purpose: We examined grafts employing for morphological analysis of early and long-term results on proximal anastomosis with the PAS-Port®.Methods: One hundred and four patients treated by OPCAB with PAS-Port® were performed postoperative MDCT. Morphological evaluation of the proximal anastomotic region was classified into three groups (A; graft was anastomosed almost perpendicularly to the aortic wall, B; graft was same type A, but subsequently curved to form an acute angle with the aortic wall, C; graft take off acute angle with the aortic wall) evaluated on planar and sagittal sections.Results: One hundred twenty-six PAS-Port® were used. Patency rate was 99.0% at discharge, 94.7% at 1 year, and no blockages were detected thereafter in patients examined. The morphology rate was A 50.6%, B 15.3% and C 34.1% on planar sections, and A 58.8%, B 10.6% and C 30.6% on sagittal sections.Conclusion: The morphological evaluation of grafts revealed the degree of freedom in graft design to be relatively high and long-term patency posed no particular problem even if the layout of the proximal anastomotic region involved a relatively acute angle. The PAS-Port® was considered to be a highly reliable device which performed appropriate proximal anastomosis and improved the patency of vein grafting to the aortic wall.

Information related to the author
Previous article Next article

Altmetrics
Recently visited articles
feedback
Top