Biological and Pharmaceutical Bulletin
Online ISSN : 1347-5215
Print ISSN : 0918-6158
ISSN-L : 0918-6158
Notes
High-Performance Liquid Chromatographic Assay of Clonazepam in Human Plasma Using a Non-porous Silica Column
Mitsuhiro NakamuraKana FukawaTadashi SugiyamaYoshihiro Katagiri
著者情報
ジャーナル フリー

2004 年 27 巻 6 号 p. 893-895

詳細
抄録

A new HPLC method has been developed for measuring clonazepam (CZP) in plasma, using a reversed-phase non-porous silica column packed with 2 μm particles. CZP in plasma was first purified with a column extraction technique and injected onto a non-porous silica column. The calibration curve was linear from 5—200 ng/ml. The recoveries of CZP added to plasma were more than 94.0%, with a coefficient of variation in the range of 5.1—13.8%. We developed a rapid routine method using a non-porous silica column that was accurate and improved solvent consumption in the measurement of CZP.

著者関連情報
© 2004 The Pharmaceutical Society of Japan
前の記事 次の記事
feedback
Top