Experimental Investigation
Effectiveness of Statin-Eluting Stent on Early Inflammatory Response and Neointimal Thickness in a Porcine Coronary Model
Katsumi MiyauchiTakatoshi KasaiTakayuki YokayamaKouichiro AiharaTakeshi KurataKan KajimotoShinya OkazakiHaruo IshiyamaHiroyuki Daida
著者情報
キーワード: Angioplasty, Restenosis, Statin, Stents
ジャーナル フリー

2008 年 72 巻 5 号 p. 832-838

詳細
抄録
Background Drug-eluting stent (DES) implantation is routine during coronary revascularization because DES significantly reduce rates of restenosis and target lesion revascularization compared with bare metal stent (BMS). However, available DES have limitations, such as late thrombosis because of delayed healing with poorer endothelialization and persistent local inflammation. Statins can inhibit cell proliferation, inflammation, and restore endothelial function. The present study evaluated the ability of stent-based cerivastatin delivery to reduce stent-induced inflammatory responses and adverse effects on endothelial function, and to inhibit neointimal hyperplasia in a porcine coronary model. Methods and Results Pigs were randomized into groups in which the coronary arteries (9 pigs, 18 coronaries in each group) had either a cerivastatin-eluting stent (CES) or a BMS. All animals survived without any adverse effects. Inflammatory cell infiltration evaluated using scanning electron microscopy on day 3 after stenting was significantly decreased in the treated vessels (inflammation score: 1.15±0.12 vs 2.43±0.34, p<0.0001). At day 28, endothelial function with intracoronary infusion of bradykinin was preserved in both the CES and BMS groups. Volumetric intravascular ultrasound images revealed decreased intimal volume in the CES group (28.3±5.4 vs 75.9±4.2 mm3, p<0.0001). Histomorphometric analysis showed reduced neointimal area (1.74±0.45 vs 3.83±0.51 mm2, p<0.0001) in the CES group despite similar injury scores (1.77±0.30 vs 1.77±0.22, p=0.97). Conclusion In porcine coronary arteries CES significantly decreased neointimal hyperplasia with a decreased early inflammatory response and without endothelial dysfunction. (Circ J 2008; 72: 832 -838)
著者関連情報
© 2008 THE JAPANESE CIRCULATION SOCIETY
前の記事 次の記事
feedback
Top