Clinical Outcomes of Transcatheter Aortic Valve Implantation in Patients With Extremely Large Annulus and SAPIEN 3 Dimensions Based on Post-Procedural Computed Tomography
Masaki MiyasakaSung-Han YoonRahul P. SharmaYoshio MaenoSandhu JaideepMasataka TaguriShigeaki KatoHiroyuki KawamoriTakahiro NomuraTomoki OchiaiShadi NemanpourTarun ChakravartyMamoo NakamuraCheng WenRaj Makkar
著者情報
ジャーナル フリー HTML
電子付録

2019 年 83 巻 3 号 p. 672-680

詳細
抄録

Background: Patients with severe aortic stenosis (AS) and an extra-large annulus (ELA) area (>683 mm2) can rarely be treated by transcatheter aortic valve replacement (TAVR) because of the size limitation of the transcatheter heart valves. This study aimed to evaluate the feasibility of TAVR using a 29-mm SAPIEN3 (S3) valve in patients with ELA and S3-dimensions by post-procedural computed tomography (post-CT).

Methods and Results: We included 261 patients undergoing TAVR using a 29-mm S3: 30 patients with ELA and 231 with non-ELA were identified. S3-dimensions were evaluated at the S3 inflow and annulus level by post-CT in 129 patients. The ELA group had a greater aortic annulus area measured by pre-procedural CT (737.3±54.7 vs. 578.4±41.9 mm2, P<0.0001), higher balloon inflation volume (36 vs. 33 mL, P<0.0001), a larger S3 area at inflow by post-CT (729.6±42.2 vs. 682.2±35.0 mm2, P<0.001), and a correlation between the inflation volume and S3 area (r=0.71, P=0.0005). No differences were observed between groups in paravalvular aortic regurgitation (PAR) ≥mild (43.3% vs. 27.6%, P=0.09), PAR ≥moderate (3.3% vs. 1.3%, P=0.39) or 1-year mortality (10.0% vs. 9.1%, P=0.87).

Conclusions: TAVR using a 29-mm S3 with extra inflation of the delivery balloon can be considered as a treatment option for patients with severe AS and ELA.

著者関連情報
© 2019 THE JAPANESE CIRCULATION SOCIETY