Chemical and Pharmaceutical Bulletin
Online ISSN : 1347-5223
Print ISSN : 0009-2363
ISSN-L : 0009-2363
Regular Articles
HPLC Analysis of Orlistat and Its Application to Drug Quality Control Studies
Effat SouriHassan JalalizadehAbbas Kebriaee-ZadehBettina Zadehvakili
著者情報
ジャーナル フリー

2007 年 55 巻 2 号 p. 251-254

詳細
抄録
In this study, a high performance liquid chromatography method with UV detection was developed for determination of orlistat. The chromatographic system consisted of a Nova-Pack C18 column, an isocratic mobile phase of phosphoric acid 0.1%–acetonitrile (10 : 90, v/v) and UV detection at 205 nm. Orlistat was eluted at about 6 min with no interfering peak from excipients used for preparation of dosage form. The method was linear over the range of 10—160 μg/ml orlistat (r2>0.9999). The within-day and between-day precision values were also in the range of 0.10—0.59%. The appropriate dissolution conditions were also determined and applied to evaluate the dissolution profile of orlistat capsules. Optimal conditions were 1000 ml of 3% SLS in water as dissolution medium and paddle at 100 rotation per minute. The proposed method was applied successfully to the determination of orlistat content in capsules and in vitro dissolution studies.
著者関連情報
© 2007 The Pharmaceutical Society of Japan