Application of Nondestructive Testing in Cyclic Fatigue Evaluation of Endodontic Ni-Ti Rotary Instruments
Uei-Ming LIChow-Shing SHINWan-Hong LANChun-Pin LIN
著者情報
ジャーナル フリー

2006 年 25 巻 2 号 p. 247-252

詳細
抄録
The purpose of this study was to investigate the application of nondestructive testing in cyclic fatigue evaluation of endodontic ProFile® nickel-titanium (NiTi) rotary instruments. As-received ProFile® instruments were made to rotate freely in sloped metal blocks by a contra-angle handpiece mounted on a testing machine. Rotation was interrupted periodically, and the instrument removed and engaged onto a device to monitor its stiffness by using two strain gauges in four different directions. This monitoring method has the potential to be developed into a convenient, nondestructive turnkey system that allows in situ assessment of the integrity of NiTi instruments in the clinic. Upon fracture, which was indicated by a change in instrument stiffness, the fractured surface would be examined under a scanning electron microscope. Microscopic evaluation indicated a small area of fatigue fracture with a large area of final ductile fracture, whereby the latter was the major cyclic failure mode. Based on the results of this study, we concluded that a potential nondestructive integrity assessment method for NiTi rotary instruments was developed.
著者関連情報
© 2006 The Japanese Society for Dental Materials and Devices