Original papers
Reducing, by Pulse Width Modulation, the Curing Temperature of a Prototype High-power LED Light Curing Unit
Ta-Ko HUANGChun-Cheng HUNGChi-Cheng TSAI
著者情報
ジャーナル フリー

2006 年 25 巻 2 号 p. 309-315

詳細
抄録
Third-generation LEDs have high irradiance and efficiency, but the associated temperature rise is potentially hazardous to the pulp of teeth. We evaluated, during composite polymerization, the irradiance and temperature rise of a prototype high-power LED light curing unit (LCU) with optimal pulse width modulation (PWM) , and then compared the results with four off-the-shelf high-power LCUs. A cavity was prepared in a tooth, and a composite resin layer was applied and cured. For each LCU, the irradiance and temperature changes at the pulp-dentin junction were measured. Microhardness (Vickers hardness) of cured composite samples was measured for each LCU. Our prototype had a final temperature of 36.4±1.3°C and irradiance of 1,182±1 mW/ cm2. The unit with the highest temperature had a temperature of 48.7±1.2°C and an irradiance of 1,194±1 mW/ cm2. Based on the results of the present study, it was shown that PWM technology reduced the curing temperature while retaining the polymerization effectiveness of a high-power LED LCU.