2006 Volume 25 Issue 3 Pages 604-610
The purpose of this study was to investigate the accuracy of temporary fixation with laser welding for fixed partial dentures (FPDs). Five kinds of experimental FPD with different welding/soldering gaps were fabricated (0, 20, 50μm for welding; 300μm for soldering). Then, FPDs were temporary-fixed by laser welding or with a self-curing resin. Fixation accuracy was evaluated by the change in distance and the angular deformation between two retainers. The change in distance and the angular deformation between two retainers of the FPD without welding/ soldering gap were significantly larger than the other FPDs (p<0.05). With due consideration to the displacement of teeth or implants especially in the mesiodistal direction, and by taking into account the inevitable errors of the indirect method, it seemed reasonable to provide a welding space of approximately 20μm.