Dental Materials Journal
Online ISSN : 1881-1361
Print ISSN : 0287-4547
ISSN-L : 0287-4547

この記事には本公開記事があります。本公開記事を参照してください。
引用する場合も本公開記事を引用してください。

Cyclic fatigue resistance of OneShape and WaveOne instruments using different angles of curvature
Shaymaa E. ELSAKAAmr M. ELNAGHY
著者情報
ジャーナル フリー 早期公開

論文ID: 2014-252

この記事には本公開記事があります。
詳細
抄録
This study compared the cyclic fatigue resistance of OneShape (OS) instrument which used continuous rotation with WaveOne (WO) instrument which used reciprocating motion. OS size 25, 0.06 taper and WO primary size 25, 0.08 taper were used in simulated canals with 45°, 60°, and 90° angles of curvature until fracture. The number of cycles to fracture (NCF) was recorded to evaluate cyclic fatigue resistance. Data were statistically analyzed using two-way analysis of variance and Tukey’s HSD test. Weibull analysis was performed on NCF data. WO had a significantly higher NCF than OS for all angles of curvature (p<0.05). According to Weibull distribution, WO instrument was predicted to have a higher number of cycles to survive than OS instrument. WO instrument which used the reciprocating motion had a greater resistance to cyclic fatigue than OS which used continuous rotation movement.
著者関連情報
© 2015 The Japanese Society for Dental Materials and Devices
feedback
Top