IEICE Electronics Express
Online ISSN : 1349-2543
ISSN-L : 1349-2543
LETTER
A current-reuse MICS band CMOS RF transmitter for implantable cardioverter telemetry system
Jongsik KimHyunchol Shin
Author information
JOURNAL FREE ACCESS

2013 Volume 10 Issue 12 Pages 20130235

Details
Abstract

A MICS band CMOS RF transmitter is presented for implantable cardioverter telemtry system. Current-reuse and gain boosting techniques are employed for reducing the current consumption as well as improving the output power. Fabricated in 0.18µm CMOS, the transmitter dissipates 9.3mW from a 1.8-V supply. Measurement shows that the gain boosting capability increases the output power by 1.3-5.5dB, while the image and LO suppression is unaffected. The proposed approach can be effective for a low-power RF transmitter design in implantable telemetry devices.

Content from these authors
© 2013 by The Institute of Electronics, Information and Communication Engineers
Previous article Next article
feedback
Top