Functional Form Examination of Recalibration Technique for Long Time CTA Measurements of Experiments with a Duplex Differential Manometer
Hiroki Suzuki Keita NakamuraToshinori Kouchi
著者情報
ジャーナル オープンアクセス

2025 年 3 巻 p. 1-17

詳細
抄録
The purpose of this study is to reduce the uncertainty of reference velocity measure-ments in low velocity ranges using a duplex differential manometer and to evaluate the need for an additional term in King's law to approximate a calibration curve of a constant-temperature hot-wire anemometer. Based on these findings, the study further investigates a hot-wire recalibration method that allows long-term measurements, focusing on the calibration curve function. While previous studies have relied on the output voltage itself for recalibration, this study adopts a method based on the squared output voltage according to King's law. The accuracy of these two hot-wire recalibration methods is compared, and the results indicate that they are equivalent. In addition, to validate the feasibility of long-term hot-wire measurements, mean flow quantities were measured over nearly six hours using a Pitot tube and a hot-wire anemometer. The uncertainties of all measurements were confirmed to be less than 1%. These results underscore the reliability of the hot-wire recalibration method as a calibration technique and pave the way for long-term hot-wire measurements.
著者関連情報
© 2025 Author, Flow Energy Association

この記事はクリエイティブ・コモンズ [表示 - 非営利 - 改変禁止 4.0 国際]ライセンスの下に提供されています。