電気学会論文誌E(センサ・マイクロマシン部門誌)
Online ISSN : 1347-5525
Print ISSN : 1341-8939
ISSN-L : 1341-8939
特集論文
小型収束超音波トランスデューサを用いた内視鏡的治療デバイス
安居 晃啓芳賀 洋一陳 俊傑伊関 洋江刺 正喜和田 仁
著者情報
ジャーナル フリー

2007 年 127 巻 2 号 p. 69-74

詳細
抄録
In this research, an ultrasonic probe (5.5 mm in diameter), which has a concave PZT transducer at its tip, was fabricated for ultrasonic treatments such as sonoporation and sonodynamic therapy in the human body using a catheter and/or endoscope. Ultrasound has the potential to enhance cytotoxicity of drugs such as porphyrins, a process referred to as sonodynamic therapy, and also to deliver macromolecules such as plasmid DNA, a process referred to as sonoporation. The fabricated probe was then experimentally characterized by measuring the acoustic intensity distribution around the focal point, using a PVDF needle-type ultrasonic hydrophone. When the PZT transducer was driven by a 120 Volts peak-to-peak AC signal at 1.83 MHz, the ultrasound output was successfully focused at the focal point, with a peak intensity of 24.9 W/cm2 (0.87 MPa). Using the fabricated probe, cultured Chinese Hamster Ovary (CHO) cells were exposed to ultrasound (1.83 MHz, continuous wave, peak acoustic pressure of 0.5 MPa) for 2 s in the presence of microbubbles MB-3 and Green Fluorescent Protein (GFP) plasmid DNA. As a result of sonication, the expression of GFP was observed in CHO cells.
著者関連情報
© 電気学会 2007
前の記事 次の記事
feedback
Top