電気学会論文誌E(センサ・マイクロマシン部門誌)
Online ISSN : 1347-5525
Print ISSN : 1341-8939
ISSN-L : 1341-8939
論文
生体信号モニタリング用スマートシャツに適した伸縮性ドライ電極の検討
天野 祐作多田 泰徳佐藤 友信齋藤 繁井上 雅博
著者情報
ジャーナル フリー

2017 年 137 巻 12 号 p. 426-431

詳細
抄録

Dry electrodes of a smart shirt for measuring biological signals were prepared by the additive printing method using a stretchable conductive paste with Ag fillers loading. Interfacial impedance between the electrode and skin depended on the contact pressure of electrode and the moisture of stratum corneum. When moisture of stratum corneum was more than 30%, ECG could be measured using the stretchable electrodes printed on a compression shirt. However, noise signal in the ECG signal increased when the moisture of stratum corneum was less than 30%. To solve this problem, we propose to use additional electrodes made of an electrically conductive kneaded rubber.

著者関連情報
© 2017 電気学会