電気学会論文誌E(センサ・マイクロマシン部門誌)
Online ISSN : 1347-5525
Print ISSN : 1341-8939
ISSN-L : 1341-8939
論文
マイクロ流体デバイス内の電流密度解析に基づいた経上皮電気抵抗測定用電極の設計法
宮崎 貴史平井 義和亀井 謙一郎土屋 智由田畑 修
著者情報
ジャーナル 認証あり

2020 年 140 巻 10 号 p. 285-292

詳細
抄録

Trans-epithelial electrical resistance (TEER) is a widely used non-invasive and label-free measurement to evaluate the barrier function of mono-layered cells in real-time. However, TEER values depend on distribution of electric current densities generated by a combination of shapes and arrangements of plane electrodes for performing TEER, but until now, there is no systematical analysis of their effects on TEER values, causing less reproducibility. Here, we report a new strategy to design the shapes and arrangements of electrodes to perform TEER in a microfluidic device. To investigate the effects of the electrodes design on TEER values, we analyze the distribution of electric current densities using finite element method and evaluate their uniformity of sensitivity. Our strategy allowed designing several electrode patterns of electrodes combinations, capable of simultaneous cell observation with a microscope and TEER measurements.

著者関連情報
© 2020 電気学会
前の記事 次の記事
feedback
Top