中国・四国整形外科学会雑誌
Online ISSN : 1347-5606
Print ISSN : 0915-2695
ISSN-L : 0915-2695
原著
ポーラスコーティングの破綻によって loosening を来したセメントレス人工股関節の2例
山口 郁子森 諭史岩田 憲眞鍋 健史真柴 賛阿部 真行乗松 尋道
著者情報
ジャーナル 認証あり

2004 年 16 巻 2 号 p. 259-263

詳細
抄録
The authors report 2 cases of cementless hip prosthesis in which component loosening was cased by failure of porous coating. The first cases was a Harris Galante type stem of 17 years after surgery. Since 1 year after surgery separation of fiber mesh coating from the stem was detected on radiographs. At revision surgery, no bone ingrowth was observed on the stem. The second case was PSL socket with a beaded titanium coating. At revision surgery, a number of beads were found to have fallen from the socket surface. Some of the beads had migrated into the joint surface, and induced metallosis. The loosening of the prosthesis is attributed to failure of the porous coating. This report calls attention to the need for close follow-up observation of patients with these prostheses.
著者関連情報
© 2004 中国・四国整形外科学会