GAF Systemを用いた腹部IVR時の患者皮膚線量の検討
松下 明民勝田 稔三河村 武人若狭 弘之門久 繁文中桐 義忠
著者情報
ジャーナル フリー

2007 年 63 巻 5 号 p. 586-594

詳細
抄録

Radiation-induced skin injury caused by interventional radiology(IVR)is a deterministic effect. If exposure dose exceeds threshold dose, injuries may occur. It therefore is important to understand the maximum exposure dose in skin. The purpose of this study was to determine the maximum exposure dose and its dose distribution. Moreover, to analyze the factors from measuring the absorbed dose in the incoming radiation side, a film-type dosimeter was used. When the measured results were determined in terms of the clinical aspect, fluoroscopic time(total time)in procedure, it ranged from 3.3 to 64.0 minutes, and DSA images obtained ranged from 8 to 280 images. Absorbed dose ranged from 0.3 to 6.0 Gy, with an average dose of 3.2 Gy. It thus exceeded 2 or 3 Gy, which was the threshold dose of temporary erythematic or depilation in 10 of 14 cases. The maximum dose was 6.0 Gy for a procedure of percutaneous transhepatic obliteration. The maximum exposure dose can be determined objectively by using a film-type dosimeter. It was also possible to grasp the overall dose distribution visually