原著
外部放射線治療におけるシステマティックセットアップエラー修正法に関する検討
高倉 亨松原 佳菜溝脇 尚志中田 学矢野 慎輔岡田 孝藤本 隆広加茂前 健田中 利恵越田 吉郎菊池 雄三
著者情報
ジャーナル フリー

2008 年 64 巻 1 号 p. 35-40

詳細
抄録
Background and Purpose: We verified the propriety of our systematic error reduction strategy by means of a computer simulation based on our data of position error with a prone fixation device for prostate IMRT. Materials and Methods: Computer simulations of the off-line correction method for systematic setup errors based on the portal imaging taken on the first several days of the treatment session were performed. Using the computer simulations, an optimal number of portal images were evaluated for the SD value, from 0.5 mm to 1.5 mm at a 0.25 mm interval, and the respective required setup margins were calculated. Results: The value of systematic error was reduced as the frequency of data obtained increased. Moreover, the reduction rate was so remarkable that random error was large.