原著
Digital Mammographyにおける被曝線量低減の検討―入射線量と処理画像画質の関係―
藤村 良夫西山 光舛本 俊典河野 信吾北川 康子池田 俊貴古川 隆志石田 隆行
著者情報
ジャーナル フリー

2008 年 64 巻 2 号 p. 259-267

詳細
抄録

In digital mammograms, granularity is an important image property for the detection of microcalcifications and masses. Therefore, we investigated the relationship between the conditions of various exposure doses and the detectability of RMI156 phantom images with and without image processing for the reduction of exposure dose. The images are processed with Gaussian filter and unsharp-masking filters to evaluate the effects on image properties by using the digital Wiener spectrum (WS) presampled modulation transfer function (MTF). In addition, observer performance tests for the detectability of microcalcifications and masses are performed. With Gaussian filtering, the WS value decreased to 50% at 2.0 cycles/mm and the detectability score of masses increased 80% and 12%, on 1.34 mGy and 2.62 mGy, respectively (p<0.05). With unsharp-masking (7×7 pixels), the MTF value increased to 126% at 2.0 cycles/mm, and the detectability of microcalcification to 32% and 5%, on 1.34 mGy and 5.28 mGy, respectively (p<0.05) compared with the original image. The optimal dose of simulated lesions with unsharp masking became 5.25 mGy. The unsharp masking could reduce 37% of the exposure dose without a loss of detectability of microcalcifications and masses.

著者関連情報
© 2008 公益社団法人 日本放射線技術学会