日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
速報
MRIマグネットが刺青シールおよびアイメイクに及ぼす力学的作用の検討
森下 雄太宮地 利明上田 丞政清水 満濱口 隆史藤原 康博林 弘之
著者情報
ジャーナル フリー

2008 年 64 巻 5 号 p. 587-590

詳細
抄録

The purpose of our study was to assess the mechanical effect on tattoo seals and eye makeup caused by a spatial magnetic gradient in the magnetic resonance imaging (MRI) system. Seven kinds of tattoo seals and three kinds of eye makeup, i.e., mascara, eye shadow, and eyeliner were used. On a 3.0-Tesla MRI, we determined these deflection angles according to a method established by the American Society for Testing and Materials (ASTM) at the position that produced the greatest magnetically induced deflection. Eighty-five percent of the tattoo seals showed deflection angles greater than 45 degrees of the ASTM guidelines, and the mascara and eye shadow showed over 40 degrees. This was because these contained ferromagnetic pigments such as an iron oxide, but those translational forces were very small owing to slight mass. However, it is desirable that these should be removed before MRI examination to prevent secondary problems.

著者関連情報
© 2008 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top