日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
ノート
ガントリに傾斜を付け撮影することで生じる画質への影響と補正-新しく考案した球体ファントムを用いての画質の補正-
古屋 研中村 公二鬼塚 牧子渡辺 梨恵長谷部 裕介野方 容子岡田 大樹荒木 力
著者情報
ジャーナル フリー

2008 年 64 巻 8 号 p. 960-970

詳細
抄録
The influence of gantry inclination on image quality was examined in the Y and Z directions and an increase in the volume of a scan part. FWHM (full width at half maximum) in the Y direction decreased at the rate of 2.53%, which was caused by leaving the center of the gantry, and spatial resolution was improved. FWHM (y) in the Z direction increased at a rate of y=2.4× (1/cosθ)+0.91 which was caused by an increase in the gantry angle of inclination, and spatial resolution was declined. The relation between FWHM (y) and pitch (x) was y=0.19x+3.12. The increase in FWHM caused by the increase in the gantry angle of inclination was revised with pitch from two formulae, and it became constant. Noise was corrected in the same way, and it became constant. The increase in the noise that occurs due to the increase in the volume of the body caused by the gantry tilt angle was corrected with the tube current, and applied to AEC (automatic exposure control). The influence of gantry inclination on image quality was adjusted, and image quality was improved.
著者関連情報
© 2008 公益社団法人 日本放射線技術学会
前の記事 次の記事