X線検出器(イメージングプレート)の放射能汚染に対する 効果的な除染方法の提案
西原 貞光林 裕晃
著者情報
ジャーナル フリー

2011 年 67 巻 8 号 p. 912-915

詳細
抄録
Several hospitals have been observing black spots in medical images, and the radioisotopes (RIs) that cause the spots needs to be removed from the X-ray receptors. Our purpose is to show a flowchart for finding out under which conditions an imaging plate (IP) and other parts (for example, the cassette) are contaminated by RIs and to propose an effective method to remove them. The procedure follows. (1) Is RI activity low? (2) Are the surfaces of other parts contaminated? (3) Is the surface of the IP contaminated? (4) Are the insides of the other parts contaminated? To remove the adhered RIs, we applied a wipe test method using a wet type of chemical wiper. A certain hospital that observed black spots experimented with this method. As a result, the contaminated condition of the X-ray receptor was identified. In addition, we were able to remove the RIs from the IP. Therefore, our procedure is very effective for decontaminating adhered RIs from receptors.