日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
ノート
SPECT 収集角度と収集軌道が画像に与える影響について—心筋ディジタルファントムスタディ—
飯塚 一則山木 範泰久木 裕也滝沢 秀喜
著者情報
ジャーナル フリー

2015 年 71 巻 6 号 p. 520-526

詳細
抄録
Purpose: We evaluated the effect that collection angle and collection orbit condition gave to an image quantitatively by simulating the single photon emission computed tomography (SPECT) system. Method: Using the Software Package of the Nuclear Medicine Data Processor for Research, we performed making of the myocardial digital phantom, three ways of different simulation of the collection angle and collection orbit, and making of reference of the uniform picture element level. We calculated NMSE for uniformity evaluation and calculated myocardial thickness full width at half maximum (FWHM) for a spatial resolution evaluation. Results: 360 degrees circular orbit collection had best uniformity. 180 degrees noncircular orbit collection had best spatial resolution. Conclusion: By using the digital phantom, we focused on only collection angle and collection orbit condition, and focused on two indexes of the uniformity and the spatial resolution and were able to show a quantitative index.
著者関連情報
© 2015 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top