異なる線量正規化法による電子線モンテカルロ計算と従来の手計算法とのモニタユニット比較
塩路 真紀中田 学藤本 隆広中村 光宏石原 佳知
著者情報
ジャーナル フリー

2019 年 75 巻 8 号 p. 755-764

詳細
抄録

The purpose of this study was to evaluate the discrepancy between the monitor unit (MU) calculated by different dose normalization methods in the electron Monte Carlo (eMC) algorithm and the conventional manual MU. In the water phantom condition, the manual MU obtained from the measured output factor was compared with the calculated MU by the eMC algorithm, using 24 different irradiation field shapes and several different energies of electron beam. In the breast boost condition, calculated MUs by both calculation methods were evaluated for 45 cases. As a result, the MUs computed by the eMC algorithm in the water phantom varied according to the dose normalization methods, and the mean±standard deviation of the difference between the manual and calculated MU were 1.1±1.4%, 0.0±1.0% and 0.4±1.2% in peak depth normalization (PN), no plan normalization (NPN) and 100% at body maximum (100%BM), respectively. In breast-boost cases, the MU difference between the manual and the calculated MU were 6.1±3.7%, 3.4±2.8% and 1.1±2.9% in PN, NPN and 100%BM, respectively. We revealed that the resultant MU calculated by eMC algorithm was dependent on the dose normalization method and the averaged differences exceeded 6% in PN, especially in breast boost condition. When using the eMC in the breast boost condition, it is desirable to select an appropriate dose normalization method according to dose prescription policies at each facility.