The Journal of Medical Investigation
Online ISSN : 1349-6867
Print ISSN : 1343-1420
ISSN-L : 1343-1420
Mismatch between Augmented Reality Navigation Images and Actual Location of a Cauda Equina Tumor:A Case Report
Shutaro FujimotoFumitake TezukaTakahiro OgawaShunsuke TamakiKosuke SugiuraMakoto TakeuchiMasatoshi MorimotoKazuta YamashitaKoichi Sairyo
著者情報
ジャーナル フリー

2024 年 71 巻 1.2 号 p. 174-176

詳細
抄録

Background:Augmented reality navigation is the one of the navigation technologies that allows computer-generated virtual images to be projected onto a real-world environment. Augmented reality navigation can be used in spinal tumor surgery. However, it is unknown if there are any pitfalls when using this technique. Case presentation:The patient in this report underwent complete resection of a cauda equina tumor at the L2-L3 level using microscope-based augmented reality navigation. Although the registration error of navigation was <1 mm, we found a discrepancy between the augmented reality navigation images and the actual location of the tumor, which we have called “navigation mismatch”. This mismatch, which was caused by the mobility of the spinal tumor in the dura mater, seems to be one of the pitfalls of augmented reality navigation for spinal tumors. Conclusions:Combined use of intraoperative ultrasound and augmented reality navigation seems advisable in such cases. J. Med. Invest. 71 : 174-176, February, 2024

著者関連情報
© 2024 by The University of Tokushima Faculty of Medicine
前の記事 次の記事
feedback
Top