Accuracy of intraoral optical scan versus stereophotogrammetry for complete-arch digital implant impression: An in vitro study
Alessandro PozziEnrico AgliardiFabrizio LioKatalin NagyAlessandra NardiLorenzo Arcuri
著者情報
ジャーナル オープンアクセス 早期公開

論文ID: JPR_D_22_00251

この記事には本公開記事があります。
詳細
抄録

Purpose: To assess and compare the accuracies of intraoral scanners (IOS) and stereophotogrammetry (SPG) devices for complete-arch digital implant impressions.

Methods: A 4-analog model was digitized using a desk scanner to obtain a reference file. Thirty test scans were conducted using the investigated IOS device, while an additional 30 scans were performed using the SPG device. Using the best-fit algorithm, the resulting 60 test files were aligned with the reference file. Linear (ΔX, ΔY, and ΔZ-axis) and angular deviations (ΔANGLE) were evaluated. Three-dimensional (3D) deviation was calculated based on the Euclidean distance (ΔEUC). The analysis was stratified according to the scanning device and implant position. Fisher’s F and t-tests were used to compare the variances and expected values of the two scanning systems.

Results: IOS expressed a higher 3D (ΔEUC) mean deviation than SPG (52.8 µm vs. 33.4 µm, P < 0.0001), with extreme measurements up to 181.9 µm. A significantly higher standard deviation (SD) was associated with IOS (37.1 µm vs. 17.7 µm, P < 0.0001). Considering angular deviations, the IOS showed slightly higher angular mean deviations (ΔANGLE) than the SPG (0.28° vs. 0.24°, P = 0.0022), with extreme measurements of up to 0.73°. The SPG SD values were significantly lower than the IOS SD values (0.14° vs. 0.04°, P < 0.0001).

Conclusions: The SPG showed significantly higher 3D and angular accuracies for complete arch implant impressions, with consistent repeatability. IOS scanning revealed significantly higher extreme deviations exceeding the acceptable threshold value. Despite study limitations, SPG appears more feasible than IOS for complete-arch digital implant impressions.

著者関連情報
© 2023 Japan Prosthodontic Society

This is an open-access article distributed under the terms of Creative Commons Attribution-NonCommercial License 4.0 (CC BYNC 4.0), which allows users to distribute and copy the material in any format as long as credit is given to the Japan Prosthodontic Society. It should be noted however, that the material cannot be used for commercial purposes.