Jinko Zoki
Online ISSN : 1883-6097
Print ISSN : 0300-0818
ISSN-L : 0300-0818
Development of Handytype Plasms Exchange System Plasma Pheresis has been applied to many kinds of immunological diseases
H. KAWAGUCHIT. UCHIMAT. WATANABEK. FUKUIO. OTSUBO
Author information
JOURNAL FREE ACCESS

1983 Volume 12 Issue 1 Pages 279-282

Details
Abstract
We developed the new system with our idea that is to controll the inlet pressures of double tube pump to drain and return blood by automatic pressure controller, improving our “Handy type Plasma Exchange Hemofiltration System in 1979”. It is 250mm Width, 350mm Hight, 366mm Depth and about 15Kg Weight. It contains new type of blood pump, heparin infuser, T. M. P. and Venous pressure monitors with upper and lower limit alarms, bubble detector. Results of in vitro and in vivo examinations are satisfied with safty use and various applications. Deviation between substitution and filtrate volume is -2.93% averaged 3 times in vivo examinations (filtrate: 2000ml).
Content from these authors
© The Japanese Society for Artificial Organs
Previous article Next article
feedback
Top