血糖制御のためのopen-loop control systemの開発とその妥当性の検討
七里 元亮河盛 隆造鮴谷 佳和新居 貞雄直 克則阿部 裕
著者情報
ジャーナル フリー

1980 年 9 巻 6 号 p. 1086-1089

詳細
抄録
In order to develop an optimum program for the subcutaneous open-loop insulin infusion which can attain the physiological meal-related plasma insulin profile, the kinetics of subcutaneously administered insulin was studied by two-compartment analysis. An exponentially decreasing insulin infusion pattern was determined by mathematically deconvoluting an ideal insulin curve. Furthermore, a new insulin infusion system which cosisted of portable and preprogrammable infusion pump was developed to enable the abovementioned program. Thus, these new program and apparatus succeeded in obtaining an expected plasma insulin profile as well as glycemic normalization and proved to be feasible and promising for the long-term glycemic and insulin normalization.