Development and Performance Evaluation of Dilution System for Diagnosis Systems with Saliva
Yuji JouKeisuke MORISHIMAYuji FURUKAWA
著者情報
会議録・要旨集 フリー

詳細
抄録

In order to use saliva for human stress diagnosis, ELISA (Enzyme-Linked ImmunoSorbent Assay) is most promising, when saliva must be diluted in purified water from about 101 to 104 times corresponding to the applied stress markers. A new dilution system with 4 diluting ladders in which each step has 10 times dilutive capability is developed, hence, can realize 101 to 104 step-likely. Its performance was tested by having mixed a fluorescent material in pure water and checked its change of intensity, and it is verified that +-20% error for the 104 time dilution target is kept.