ロボティクス・メカトロニクス講演会講演概要集
Online ISSN : 2424-3124
セッションID: 2P2-20a3
会議情報

低ノイズなMRI画像計測のための二重巻きマイクロコイル
深堀 俊土肥 徹次
著者情報
キーワード: Micro coil, MEMS, Double winding, MRI, Low noise
会議録・要旨集 フリー

詳細
抄録

This paper reports on a double winding micro coil as a receiver for MRI (Magnetic Resonance Imaging). A double winding micro coil was fabricated by assembling 3D printed structures and planar coil wirings fabricated by MEMS (Micro Electro Mechanical Systems) technology. Since a cross sectional area of wires of double winding coil is two times larger than single winding coil, this double winding micro coil has low resistance. The diameter of the inner coil, outer coil, and length of the coil were 4.5 mm, 6.5 mm, and 10 mm, respectively. The number of turns of micro coils were 8 and 10. The resisitances and SNRs (Signal to Noise Ratio) of 8-turn and 10-turn coils were 0.239 Ω and 0.424 Ω and 114 and 88, respectively. Since a double winding micro coil has high performance, we acquired a clear MRI image of a small berry of Liriope muscari.

著者関連情報
© 2016 一般社団法人 日本機械学会
前の記事 次の記事
feedback
Top