Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Case Reports
Blade Crossing of a Pure Titanium Clip Applied to a Cerebral Aneurysm
—Case Report—
Yutaka HIRASHIMAMasanori KURIMOTOMichiya KUBOShunro ENDO
著者情報
ジャーナル オープンアクセス

2002 年 42 巻 3 号 p. 123-124

詳細
抄録
Titanium aneurysm clips are superior to cobalt clips in radiological features, biocompatibility, and hypoallergenicity, but the mechanical characteristics are less advantageous. A 56-year-old man was referred to our department due to dizziness and slurred speech. Cerebral angiography unexpectedly disclosed an aneurysm of the right middle cerebral artery. The patient underwent right frontotemporal craniotomy and neck clipping of the aneurysm with a commercially pure titanium clip. After application of the clip, the clip blades crossed and nearly sheared the wall of the aneurysm. The mechanical characteristics of titanium aneurysm clips require careful consideration because of possible clip malfunction.
著者関連情報
© 2002 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/