Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Case Report
Tied Pipeline: A Case of Rare Complication
Shigeru MIYACHIHiroyuki OHNISHIRyo HIRAMATSURyokichi YAGIToshihiko KUROIWA
著者情報
キーワード: aneurysm, flow diverter, complication
ジャーナル オープンアクセス

2018 年 58 巻 5 号 p. 219-224

詳細
抄録

A 48-year-old female with a symptomatic giant carotid cavernous aneurysm underwent endovascular treatment with multiple Pipeline Flex embolic devices (PED). The delivery catheter had to take a complicated course of multiple turns to cross the aneurysm, and its loop was tied in the aneurysm. When the catheter was to be withdrawn, it was apparent that the tied catheter had made a tight knot that bound the tail of the previous PED together. We successfully retrieved all of the system including the tied PED, and we used telescoping stents with five PEDs in the next operation. Although this is a rare complication, it is worth noting and checking to make sure that there is no knot before deploying the stent.

著者関連情報
© 2018 by The Japan Neurosurgical Society

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/
前の記事 次の記事
feedback
Top