Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Technical Notes
The Usefulness of the 3.1-mm-diameter 4K Rigid Endoscope for Intracerebral Hematoma Evacuation
Hiroto OKUBATAMasaaki TANIGUCHINaoki IRIZATOHidekazu NAKATATomofumi TAKENAKAShogo FUKUYAKatsunori ASAIAkihiro TATEISHIKazumi YAMAMOTOAkatsuki WAKAYAMA
著者情報
ジャーナル オープンアクセス
電子付録

2024 年 64 巻 7 号 p. 283-288

詳細
抄録

The indication for surgical intervention in spontaneous intracerebral hemorrhage remains controversial. Although many clinical trials have failed to demonstrate its efficacy over medical treatment, less invasive endoscopic treatment is expected to demonstrate its superiority. A novel endoscopic system for hematoma removal consisting of a 3.1-mm-diameter 4K high-resolution rigid endoscope was used.

The system was used in eight cases of spontaneous intracerebral hemorrhage. It provided improved maneuverability of the surgical instrument while maintaining satisfactory image quality. The surgical goal was achieved in all cases without any complications, including perioperative rebleeding.

Endoscopic hematoma removal using the 3.1 mm high-resolution endoscope is an alternative minimally invasive approach to spontaneous intracerebral hemorrhage with improved reliability.