Neurologia medico-chirurgica
Online ISSN : 1349-8029
Print ISSN : 0470-8105
ISSN-L : 0470-8105
Peltier stackを用いたthermal diffusion法による連続的脳血流測定の試み
甲州 啓二広田 茂園部 真高橋 慎一郎高久 晃斉藤 建夫牛島 豊彦
著者情報
ジャーナル フリー

1987 年 27 巻 8 号 p. 724-728

詳細
抄録
The authors report an experimental trial to record cerebral blood flow (CBF) continuously by means of a thermal diffusion method using a Peltier stack. The authors performed the experiments after Carter's method and confirmed that a stable recording was possible. To calibrate the flow probe, they used CBF values obtained by the electrolytic method. The authors proved theoretically and experimentally that calibration of the flow probe is possible without knowing the CBF values at the time of cardiac arrest.
The thermal diffusion method using a Peltier stack is a very useful method for monitoring CBF continuously, especially during neurosurgical operations. Many problems of cerebral hemodynamics should be solved by measuring CBF values by this method.