日本PDA学術誌 GMPとバリデーション
Online ISSN : 1881-1728
Print ISSN : 1344-4891
ISSN-L : 1344-4891
寄書
異なるパーティクルカウンタの計算値の違いについて
奥井 敬造
著者情報
ジャーナル フリー

2000 年 2 巻 1 号 p. 49-52

詳細
抄録
  This paper describes the effect of allowance for basic specifications of single particle counters on its counting results. Even particle counters meet requirement of Japanese Industrial Standard JIS-B-9921 (Light Scattering Automatic Particle Counter), they may show more than 40% differences in counting results. And we found that size resolution which is not specified in JIS-9921 is also important factor to get more accurate data from comparison test.
著者関連情報
© 2000 日本PDA製薬学会
前の記事
feedback