臨床血液
Online ISSN : 1882-0824
Print ISSN : 0485-1439
ISSN-L : 0485-1439
臨床研究
人内因子結合57Co-B12および遊離58Co-B12の同時投与法によるB12吸収試験について
中島 隆阿部 帥岡村 裕喜小宮 正文
著者情報
ジャーナル 認証あり

1974 年 15 巻 3 号 p. 291-297

詳細
抄録
A urinary B12 excretion test by dual isotope technique was performed in 22 patients including 7 with pernicious anemia, 3 with total gastrectomy and 2 with chronic renal failure. The urinary excretion rate of 58Co-B12 was 21.3±3.7% in control group and 4.1±2.6% in both patients with pernicious anemia and those with total gastrectomy. The ratio of 57Co-B12 to 58Co-B12 was significantly high (p<0.01), ranging 1.4 to 9.4 in the patients with IF deficiency. In this group, percent excretion of 58Co-B12 was greater than that of free B12 only when administrated to the same patient.
Free IF was not found in the solution of IF bound 57Co-B12 and 8 to 15% of 57Co-B12 was proved free from IF by coated charcoal method. It was showed that the exchange of 57Co-B12 on IF with free B12 occurred depending on pH and temperature in vitro. Therefore, the higher urinary excretion of the free form of B12 in the dual isotopic method was probably due to the exchange of two forms of B12 in vivo.
The radioactivity of 58Co-B12 in 2 ml of urine is too low to be measured accurately. This was settled by counting 58Co-B12 adsorbed on uncoated charcoal from 20 ml of urine.
著者関連情報
© 1974 一般社団法人 日本血液学会
前の記事 次の記事
feedback
Top