SICE Annual Conference Program and Abstracts
SICE Annual Conference 2002
会議情報

Uncertainty Evaluation on the Freezing Point of Silver for Thermocouple Calibration
Haruhiko NumajiriHideki OguraMasaya IzuchiMasaru Arai
著者情報
会議録・要旨集 フリー

p. 225

詳細
抄録
The freezing point of silver is realized using an open type cell. The uncertainty of the thermocouple calibration at this temperature was evaluated using Pt/Pd thermocouples. The sources of uncertainty estimated were stability and repeatability of the plateau, temperature distribution in the cell, impurity effect of the metal, surroundings pressure correction and hydrostatic pressure correction. The combined standard uncertainty is 7, 8mK.
著者関連情報
© 2002 SICE
前の記事 次の記事
feedback
Top