Bioscience, Biotechnology, and Biochemistry
Online ISSN : 1347-6947
Print ISSN : 0916-8451
Analytical Chemistry Note
Immunoassay Using Microfluid Filters Constructed by Deep X-Ray Lithography
Katsuhiro MATSUIIsao KAWAJIYuichi UTSUMIYoshiaki UKITAToshifumi ASANOMasahiro TAKEODai-ichiro KATOSeiji NEGORO
著者情報
ジャーナル フリー

2007 年 71 巻 12 号 p. 3098-3101

詳細
抄録
Microfluid filters were fabricated, which possessed 2,100 cylindrical through-bores (φ40 μm) in 200 μm-thickness polymethylmethacrylate (PMMA) sheets (φ3 mm), by deep X-ray lithography using synchrotron radiation. To evaluate the microfluid filters as a device for an immunoassay, we bound the goat anti-mouse immunogloblin G (IgG) antibody to the surface of the filters, and set the filters between reaction vessels stacked vertically in a microreactor. An enzyme-linked immunosorbent assay (ELISA) of mouse IgG using the goat anti-mouse IgG/horseradish-peroxidase (HRP) conjugate indicated that mouse IgG could be quantitatively detected in the range of 0–100 ng/ml, demonstrating the applicability of vertical microfluidic operation to the immunoassay.
著者関連情報

この記事は最新の被引用情報を取得できません。