Experimental Investigation
Inflammation and Delayed Endothelization With Overlapping Drug-Eluting Stents in a Porcine Model of In-Stent Restenosis
Sang Yup Lim, Myung Ho Jeong, Soon Jun Hong, Do Sun Lim, Jae Youn Moon, Young Joon Hong, Ju Han Kim, Youngkeun Ahn, Jung Chaee Kang
著者情報
ジャーナル フリー

2008 年 72 巻 3 号 p. 463-468

詳細
抄録
Background This study evaluated the inflammatory reaction at the site of overlapping drug-eluting stents (DES) in a porcine model of in-stent restenosis. Methods and Results Twenty bare metal stents (BMS) (group I; n=10), 20 sirolimus-eluting stents (SES) (group II: n=10), 20 paclitaxel-eluting stent (PES) (group III: n=10), and 10 PES and 10 SES (group IV: n=10) were overlapped in the left anterior descending coronary arteries of 40 pigs. Follow-up coronary angiography and histopathology were performed at 4 weeks after stenting. For the overlapped segments, the minimal luminal diameter at 4 weeks was smaller in group I than in the other groups (1.78±0.13 mm, 2.79±0.09 mm, 2.90±0.04 mm, 2.80±0.07 mm, respectively; p<0.001), and the neointimal area (5.51±0.58 mm2, 2.38±0.53 mm2, 2.07±0.37 mm2, 2.39±0.58 mm2, respectively; p<0.001) and area stenosis (68.74±4.02%, 27.79±4.73%, 23.66±3.24%, 27.63±4.07%, respectively; p<0.001) were higher in group I than in the other groups; however, the inflammatory score was higher in group III than in the other groups (1.80±0.42, 2.10±0.32, 2.90±0.31, 2.50±0.52, respectively; p<0.001) and the endothelization score was lower in group III than in the other groups (2.80±0.42, 2.30±0.67, 1.30±0.48, 2.10±0.74, respectively; p<0.001). Conclusion Compared with BMS, DES inhibit neointimal hyperplasia, but inflammation and poor endothelization occur at the site of overlapping stents. (Circ J 2008; 72: 463 - 468)
著者関連情報
© 2008 THE JAPANESE CIRCULATION SOCIETY
前の記事 次の記事
feedback
Top