Circulation Journal
Online ISSN : 1347-4820
Print ISSN : 1346-9843
ISSN-L : 1346-9843
Cardiovascular Surgery
Change of Coronary Flow by Continuous-Flow Left Ventricular Assist Device With Cardiac Beat Synchronizing System (Native Heart Load Control System) in Acute Ischemic Heart Failure Model
Akihide UmekiTakashi NishimuraMasahiko AndoYoshiaki TakewaKenji YamazakiShunei KyoMinoru OnoTomonori TsukiyaToshihide MizunoYoshiyuki TaenakaEisuke Tatsumi
著者情報
ジャーナル フリー

2013 年 77 巻 4 号 p. 995-1000

詳細
抄録
Background: A novel control system for the EVAHEART left ventricular assist device (LVAD), known as the Native Heart Load Control System (NHLCS), can change the device’s rotational speed (RS) in synchrony with the heartbeat. The system enhanced coronary flow (CoF) with the counter-pulse mode in normal goats’ hearts, so we examined the change in CoF in goats with acute ischemic heart failure (HF). Methods and Results: We studied 14 goats (56.1±6.9kg) with acute ischemic HF induced by coronary microsphere embolization. We installed EVAHEART and ran the device in 4 modes [continuous support, circuit-clamp, counter-pulse (raise RS in diastole), and co-pulse (raise RS in systole)] with 50% or 100% bypass in each mode. In comparison with the circuit-clamp mode, CoF was 121.0±14.1% in the counter-pulse mode and 102.9±7.9% in the co-pulse mode, whereas it was 113.5±10.6% in the continuous mode, with 100% bypass (P<0.05). The same difference was confirmed with 50% bypass. The results indicated that a LVAD in an acute ischemic heart enhanced CoF, and that CoF was greater in the counter-pulse mode and smaller in the co-pulse mode relative to the continuous mode. Conclusions: By using NHLCS to change CoF, recovery of native heart function with a LVAD has a better prognosis.  (Circ J 2013; 77: 995–1000)
著者関連情報
© 2013 THE JAPANESE CIRCULATION SOCIETY
前の記事 次の記事
feedback
Top