Circulation Journal
Online ISSN : 1347-4820
Print ISSN : 1346-9843
ISSN-L : 1346-9843

この記事には本公開記事があります。本公開記事を参照してください。
引用する場合も本公開記事を引用してください。

Measurement of Hemodynamics During Postural Changes Using a New Wearable Cephalic Laser Blood Flowmeter
Tetsuya FujikawaOsamu TochikuboNaoki KuraTakanori KiyokuraJunichi ShimadaSatoshi Umemura
著者情報
ジャーナル フリー 早期公開

論文ID: CJ-09-0103

この記事には本公開記事があります。
詳細
抄録
Background: Patients with orthostatic hypotension have pathologic hemodynamics related to changes in body posture. A new cephalic laser blood flowmeter that can be worn on the tragus to investigate the hemodynamics upon rising from a sitting or squatting posture was developed. Methods and Results: The relationship between cephalic hemodynamics and cerebral ischemic symptoms in 63 subjects in a sitting, squatting, and standing positions using the new device was evaluated. Transient decrease in blood pressure within 15 s after rising to an erect position possibly causes dizziness, syncope, and fall. Subjects exhibiting dizziness upon standing showed a significant decrease in the cephalic blood flow (CBF) and indirect beat-to-beat systolic blood pressure, as monitored by the Finometer, and a significant correlation was observed between the drop ratio (drop value on rising/mean value in the squatting position) of CBF and that of systolic blood pressure. Conclusions: This new wearable CBF-meter is potentially useful for estimating cephalic hemodynamics and objectively diagnosing cerebral ischemic symptoms of subjects in a standing posture.
著者関連情報
© 2009 THE JAPANESE CIRCULATION SOCIETY
feedback
Top