衛生化学
Print ISSN : 0013-273X
フレームレス原子吸光法による環境試料中のベリリウムの定量
浮島 美之永井 武永野 隆夫海野 忠市渡辺 与八郎
著者情報
ジャーナル フリー

1979 年 25 巻 4 号 p. 211-215

詳細
抄録
A simple method for measuring beryllium in environmental materials was investigated by the use of flameless atomic absorption spectrophotometer equipped with a graphite tube atomizer. Nitric acid was suitable for decomposing environmental materials such as precipitator dusts which were exhausted from refuse incinerators. Furthermore, 2 to 10 N nitric acid concentration stabilized absorbance of beryllium, and eliminated background absorptions of sodium, potassium, and calcium when the charring temperature was set at 900°. Background absorption caused by less than 1000 μg/ml of aluminium, sodium silicate, manganese, magnesium, and iron could be corrected completely by using a deuterium are lamp as a background corrector. As the atomic absorption signal of beryllium was interfered by aluminium, calcium, iron, sodium silicate, and zine, which are commonly present in the dust, a method for standard addition should be used to correct these interventions. By this method, beryllium in precipitating dust can be analyzed without any separation procedure. Beryllium content in the dust of electric precipitators and multicyclones averaged 0.17 and 0.43 μg/g, respectively. Beryllium in them scarcely dissolved in water.
著者関連情報
© 社団法人 日本薬学会
前の記事 次の記事
feedback
Top