4.リング型エミッションCT(I.RI画像構成とその評価)(第37回総会)
蜂谷 武憲菅野 巖相沢 康夫小野寺 洋庄司 安明菅 幹雄佐々木 信夫羽上 栄一広瀬 佳治古賀 健一郎田中 義宗
著者情報
ジャーナル フリー

1981 年 37 巻 5 号 p. 566-573

詳細
抄録

Emission computed tomograph (ECT) consists of detector system, mechanism for scanning, electronics for data access and control, and computer system. ECT images have to be reconstructed carefully to correct for tissue absorption, to reduce scattered radiation and so on. These were described in detail. Also here mentioned basic design and performance of HEADTOME system developed in our institute as a first dual purpose emission computed tomograph for positron and single photon emission imaging. The system has shown the resolution of 10 and 11 mm FWHM at center (of field of view) with ^<99m>Tc and ^<68>Ga, respectively, and the sensitivity of 20 Kcps/μCi/ml (with 20 mm slice spacing). The HEADTOME is routinely used in clinical study.