抄録
Lead power and oxygenated lead are widely used as materials in conventional protective aprons. However, the use of lead alone limits weight reduction in an apron. Because the amount of lead can not be decreased without reducing X-ray absorption rate. In this study, we report the efficiency of a new X-ray protective apron with a composite protective sheet. The X-ray absorption rate of the composite sheet was 4.5 percent less than that of the lead sheet. However, the effective protective efficiency of the composite sheet was the same as that of the lead sheet. The uniformity of the new X-ray protective apron was good, and the differences among products were small. The new X-ray protective apron was about 30 percent lighter than a lead apron with a 0.25mm lead equivalent. The new lighter aprons are more comfortable for the radiology staff. Therefore, the use of a composite protective apron provides the advantages of weight reduction and increased comfort without reducing X-ray absorption rate.