X線吸収体と透視管電圧がimage intensifier(I.I.)の入射線量率に及ぼす影響
瀬口 繁信石川 芳信桑原 和義守田 美紀水野 昌太中村 覈郎
著者情報
ジャーナル フリー

1999 年 55 巻 7 号 p. 687-691

詳細
抄録

We examined the influence of phantoms and tube voltage in fluoroscopy on the image intensifier (I.I.) conversion factor. We used 20-cm-thick acrylic resin, 20 mm aluminum, and 1.5 mm copper, which are generally used as phantoms in the measurement of I.I. incident dose rate. We measured I.I. incident dose rate and conversion factor under conditions in which the range of tube voltage was from 60 kV to 120 kV. The result showed that the conversion factor is influenced by the type of phantom, with copper showing the highest value, aluminum second, and acrylic the smallest under the same condition of aluminum at half-value layer. It was determined that conversion factor depends on tube voltage and has peaks from 80-100 kV. The location and height of the peak are influenced by the type of phantom. Therefore, I.I. incident dose rate is influenced by both the type of phantom and tube voltage under automatic brightness control fluoroscopy. Unification of phantoms and tube voltage is necessary for long-term evaluation of I.I. incident dose rate.