日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
CCDカメラを用いたI.I.-DR装置の簡便な精度管理法
五十嵐 均白石 明久倉石 政彦
著者情報
ジャーナル フリー

2000 年 56 巻 2 号 p. 290-296

詳細
抄録

With the advancing development of CCD cameras, the I.I-DR apparatus has beenintroduced into the x-ray fluoroscopy television system.Consequently, quality control of the system has become a complicatedtask. We developed a simple, accurate method for quality control of the I.I.-DR apparatus using the CCD camera.Experiments were separately performed for the imager system [(laser imager, DDX(dynamic digital x-ray system )]and the imaging system(I.I., ND-filter, IRIS, CCDcamera).Quality control of the imager system was done by simply examining both input and output characteristics with a sliding pattern.Quality control of the imaging system was also conducted by estimating AVE(the average volume element), which was obtained using a phantom under the constant conditions.The results indicated that this simplified method is useful as a weekly quality control check of the I.I.-DR apparatus using the CCD camera.

著者関連情報
© 2000 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top