When chest radiographs are obtained using an anti-scatter grid in a mobile x-ray unit by Fuji Computed Radiography (FCR), right lung densities may differ from left lung densities owing to misalignment caused by an oblique x-ray beam. Such misalignment is considered to be caused by factors such as the shape of the bed and angle of the x-ray tube. Misalignment is especially frequent when exposure is carried out in the sitting position. Although a low-ratio grid has been reported to offer improvement, the misalignment is not completely eliminated. Therefore, we used a copper plate without a lattice feature instead of grid, and examined the results. The copper plate was found to have the ability of a grid to eliminate radiation scatter, and, furthermore, to offer the physical capabilities of a grid as determined by Carlin's new grid assessment method. Image quality was demonstrated through contrast manipulation to be equal to that obtained by grid use. Further, optimal conditions were found to be a 95 kV tube voltage and 0.3 mm thick copper plate.