フラットパネルディテクタを用いたデュアルエネルギーサブトラクション法におけるびまん性肺疾患の検出能の評価
木口 雅夫北川 康子藤岡 知加子石風呂 実山口 貴子池田 俊貴古川 隆志
著者情報
ジャーナル フリー

2007 年 63 巻 12 号 p. 1362-1369

詳細
抄録

We investigated and evaluated the detection of simulated lesions in various interstitial lung diseases using the dual-energy subtraction radiography method and flat-panel detector (FPD) images. We obtained a FPD system (GE Revolution XR/d), and employed dual-energy 60 kV and 130 kV exposure techniques. Three types of lung lesions, namely, micro-nodule, ground-glass, and honeycomb patterns were simulated with interstitial lung disease on a chest phantom. Chest images with and without simulated lesions were exposed and compared with standard images and subtraction images. We carried out evaluations with and without subtraction images and performed the analysis by using receiver operating characteristic (ROC) analysis of detection. Results showed that the detection of interstitial lung diseases was significantly improved by the use of subtraction images. The area under the ROC curve (AUC) values of micro-nodule images obtained with and without subtraction images were 0.768 and 0.963, ground-glass images 0.670 and 0.917, and honeycomb images 0.768 and 0.996, respectively. A significant difference of p<0.05 was accepted. The use of dual-energy subtraction radiography with a FPD improved diagnostic accuracy in detecting cases of multiple interstitial lung diseases and was considered useful.

著者関連情報
© 2007 公益社団法人 日本放射線技術学会